…3 ELECTRICAL INSTALLATION 3.2.4 Conduit Adapters and Cable Glands (North American – 0.5in) Alternative Alternative Face Seal Face Seal Face Seal Face Seal Ferrule Illustration Fittings vary Seal typical for slightly for (Present in O.Z. O.Z.Gedney different makes Gedney Fittings) Outer Nut Outer Nut Conduit Adapters...
Page 11
3 ELECTRICAL INSTALLATION… Caution. (Remote versions) • Remove any exposed black conductive layer from under coaxial screens. • Make connections only as shown. • Sleeve all bare wiring. • Twist RED and YELLOW cores lightly together. • Twist WHITE and GREY coaxial cables lightly together. •...
…3 ELECTRICAL INSTALLATION 3.3.2 Transmitters (All versions) Remove Protection Cover Caution. Unused cable entries must be blanked with the permanent blanking Slide plugs supplied with down the transmitters. Pull out slightly... Slacken ...and captive slide off screws Fig. 3.11 Transmitter Connection Terminal Caution.
Page 13
3 ELECTRICAL INSTALLATION… …3.3.2 Transmitters (All versions) (CD 1) Drive Connections Yellow White Coax (CD 2) Blue Inner (SIG 1) Screen (DS 1) Signal Connections Violet Grey Coax Pink Inner (SIG 2) Screen (DS 2) Ground Wire (Safety Earth) Fig. 3.13 Sensor Cable Connections at the Transmitter (Remote version) North American Wiring Practice (CD 1) Drive...
…3 ELECTRICAL INSTALLATION 3.3.3 Alternate Wiring Configuration Some later transmitters have an alternative (plug-and-socket) sensor wiring configuration (see Fig. 3.15) This connector may be either an integral part of the termination area or, alternatively, part of the CalMaster adapter board. The wiring of both these variants is the same. To wire the adaptor plug, carefully pull off the plug from the adaptor board, connect the wires (using a screwdriver with a 2.5mm blade to tighten the terminal screws) and replace the plug.
• Operation of outputs is programmable – see Configuration Manual for details. • Frequency and Alarm outputs share a common return with contact input. • External isolators not normally required, as the pulse, alarm and contact circuits are electrically separated from all other Magmaster connections. Timer MagMaster Transmitter...
3.4.5 Current Output – Fig. 3.20 and 3.21 Information. • Output is fully programmable – see Programming Guide. • Output is electrically separated from all other MagMaster connections. • External isolators are not normally required and may significantly limit accuracy if used. MagMaster Transmitter HART link IC –...
…3 ELECTRICAL INSTALLATION 3.4.6 Computer Connection – Fig. 3.22 and 3.23 Information. RS422/423 option is electrically isolated from all other MagMaster connections. MagMaster RS422 Connection APPLE Connector TERMINALS NAME (8 Pin MC) TX-SIG RX DATA - TX+SIG RX DATA +...
• DISCONNECT THE SUPPLY FROM ANY CABLES BEING TERMINATED ON THE TRANSMITTER. • Electrical installation and earthing (grounding) must be in accordance with relevant national and local standards. ABB Limited Line (L/L1) to L Transmitter Label SUPPLY 95V — 240V AC...
…3 ELECTRICAL INSTALLATION ABB Limited Positive to + Transmitter Label SUPPLY 11V — 40V DC (Max) Internal External Negative to – DC Supply >4mm (<10 A.W.G.) Copper Wire Fig. 3.25 Power Supply Connections (DC Version Transmitter) 3.4.8 Profibus Connections Refer to the separate manual (IM/MM/PBS) for details.
• For all versions supporting HART flow rate will be shown on the display as shown in see the main MagMaster manual. Fig. 4.1 or 4.2. Sequential application of the provided magnetic...
Page 22
…4 STARTUP AND OPERATION 12.328 Ltr/s >43567 MagMaster 32.8 Ltr/s >42315 Fig. 4.2 Location of Controls (Keypad Versions)
APPENDIX A – ENVIRONMENTAL PROTECTION Warning. • Potting materials are toxic – use suitable safety precautions. • Read the manufacturers instructions carefully before preparing the potting material. • The remote sensor terminal box connections must be potted immediately on completion to prevent the ingress of moisture.
Page 25
600mm as flange rating in. NPT connections 700mm 6, 10 or 16 bar Sensor cable Environmental protection ABB cable supplied as standard IP68 (NEMA6) Armored version available on request Buriable to 5m (16 ft) depth Power supply* Pressure equipment directive 97/23/EC...
Page 26
…SPECIFICATION Output/Inputs Mounting Common <21mA, 16V 45° maximum <800Hz, <35V open collector Forward square wave, or fixed pulse Reverse width <2.5s, 250mA Choice of <35V, >250mA open collector 9 alarms Contact closure or logic input Totalizer reset Dual range selection Sensor Electrodes Output hold Drive to zero...
Page 27
SPECIFICATION Sensor Specification (nominal dimensions) 15 to 2200mm (0.5 to 84 in.) ) . n ) . n ) b l * * ) * * ) * * ) * * ) * * ) * * ) * * ) * * ) * * * ) 6 .
Page 31
– Food & Beverage Repair Centre. – Manufacturing – Metals and Minerals United Kingdom – Oil, Gas & Petrochemical ABB Limited – Pulp and Paper Tel: +44 (0)1453 826661 Fax: +44 (0)1453 829671 Drives and Motors • AC and DC Drives, AC and DC Machines, AC...